It was reported this was a closure using the pre-close technique prior to an unspecified procedure.Reportedly, the footplate on three prostyle devices did not close.Force had to be used to remove all prostyle devices.The method used to achieve hemostasis was not provided.There was no reported adverse patient sequela and no reported clinically significant delay in the procedure or therapy.Additional information was received stating: device inserted per the instructions for use.On attempting to remove the device after deployment (after step 4 to close the footplate) the device felt stuck in the artery.Had to pull more forcefully than usual.The footplate was examined externally and seen to be not fully retracted.A total of 3 devices were used in this patient and the same issue happened with all of them.There was significant effect to patient as procedure was performed under local anesthetic.Marked pain when retrieving devices and very poor hemostasis at the end, requiring prolonged manual pressure and delay.No additional information was provided.
|