• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; FIBEROPTIC BRONCHOSCOPE 1.2C 2.7TP 600L

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; FIBEROPTIC BRONCHOSCOPE 1.2C 2.7TP 600L Back to Search Results
Model Number FB-8V
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2024
Event Type  malfunction  
Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cfb (coherent fiber bundle) broken.Based on the result, we concluded that it was caused due to the excessive force applied on the cfb (coherent fiber bundle).Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event Description
The time of event is not during procedure.There was no report of patient harm.Fiber image failure(broken).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
FIBEROPTIC BRONCHOSCOPE 1.2C 2.7TP 600L
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
gurvinder nanda
3 paragon drive
montvale, NJ 07645
2015712318
MDR Report Key18956877
MDR Text Key338562550
Report Number9610877-2024-52391
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
K951199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFB-8V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2024
Date Manufacturer Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-