ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; No Match Back to Search Results
Model Number 1458Q/92
Device Problems Break (1069); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2024
Event Type  malfunction  
Event Description
It was reported that the patient presented for a lead revision procedure.During the procedure, it was noted that the stylet was failing to be removed from the left ventricular (lv) lead.The lv lead was damaged when attempting to remove the stylet.The lv lead was not used.The physician elected to use another lv lead and completed the procedure.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of stylet stuck and ¿the lead broke while attempting to pull stylet out¿ were confirmed.A complete lead with stuck stylet was returned in one piece.The ptfe coating of the stuck stylet was stripped and was found bunched up with the inner coil distal to the connector pin.The connector pin with the crimp sleeve were found pulled out of the connector assembly stretching the inner coil consistent with damage due to excessive forces applied while attempting to remove the stylet from the lead during the procedure.The connector cap was intact and in place in the connector assembly.The cause of the reported events was isolated to the bunching of the stylet ptfe coating inside the inner coil at the connector region that prevented the removal of the stylet and excessive forces resulted in the connector pin with the crimp sleeve to be pulled out of the connector assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18957077
MDR Text Key338304583
Report Number2017865-2024-35636
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503358
UDI-Public05414734503358
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/92
Device Lot NumberA000145297
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA TACHY MRI LEAD; GALLANT HF ICD; MRI COND. TENDRIL LEAD
Patient Age78 YR
Patient SexFemale
Patient Weight70 KG
-
-