GYRUS ACMI, INC. GYRUS, PK-SP GENERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Model Number 744000 |
Device Problem
No Device Output (1435)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Based on the results of the investigation, a definitive root cause for the generator not recognizing the electrode was due to a faulty electrode socket.A device history review revealed no issues that could have caused or contributed to the reported issue.The subject device was manufactured in aug 2021.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was observed that during the device inspection, the generator was not recognizing plasma kinetic electrode due to intermittent failure of the output socket.There were no reports of patient involvement.
|
|