• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2024
Event Type  malfunction  
Event Description
It was reported that technical intervention was required as the device would not turn on.There was no patient involvement and no harm/adverse event reported.
 
Manufacturer Narrative
Date of event: unknown.No information has been provided to date.Initial reporter phone#: (b)(6).Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection found the device is in a good physical condition.Functional testing confirmed the complaint.Root cause was attributed a defective main board, which was replaced.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key18960550
MDR Text Key338455837
Report Number3012307300-2024-01607
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public(01)50695085820909(11)080916
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberHL-90-IT-230
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/12/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-