• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HP054
Device Problem Difficult or Delayed Activation (2577)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 3/22/2024.B3: event month and year only reported: january 2024.D4 batch #: r92w9f.Additional information was received: the device works intermittently.The generator recognize the device.The handpiece out of order for 2023 does not pass the pre-test but the second hand piece pass the pre-test.No patient consequence.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with the nose cone cracked and the mount was noted to be loose.No functional testing could be performed due to the nose cone was extremely cracked and no instrument could be attached to the handpiece.The instrument was disassembled to inspect internal components.The moisture indicator was positive.Due to the cracked nose cone, moisture entered the hand piece mid housing.It is possible that the ingress of moisture affected handpiece functionality.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.Although the analysis was unable to determine the exact root cause that lead to crack in the hand piece nose cone.
 
Event Description
It was reported that during an unknown procedure the handpiece is no longer functional (or would sometimes work), while they have between 40 and 50 uses left for each.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC SCALPEL HAND PIECE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18961559
MDR Text Key338682929
Report Number3005075853-2024-02338
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036002147
UDI-Public10705036002147
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K002906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHP054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2024
Date Manufacturer Received03/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/11/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
GENERATOR.
-
-