MPRI SPRINT QUATTRO SECURE MRI SURESCAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 694765
Device Problems Intermittent Capture (1080); Fracture (1260); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Bradycardia (1751); Shock from Patient Lead(s) (3162); Syncope/Fainting (4411)
Event Date 03/01/2024
Event Type  Injury  
Event Description
It was reported that the right ventricular (rv) lead had a low-voltage fracture resulting in inappropriate shock therapy and intermittent rv capture.It was additionally reported that the patient experience syncope and bradycardia as a result.The lead was capped and replaced when the patient was downgraded to a pacemaker system. no further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE MRI SURESCAN
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18961585
MDR Text Key338384515
Report Number2649622-2024-07978
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P920015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/27/2005
Device Model Number694765
Device Catalogue Number694765
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/01/2024
Date Device Manufactured03/27/2003
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DVFB1D1 ICD
Patient Outcome(s) Life Threatening; Required Intervention; Hospitalization;
Patient Age84 YR
Patient SexFemale
-
-