THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 102956
Device Problems Infusion or Flow Problem (2964); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2024
Event Type  malfunction  
Event Description
It was reported that the centrimag motor failed when it was not in use on a patient.The clinician reported that they were setting up an extracorporeal membrane oxygenation circuit and testing the device and got error s3, motor alarm m4, and the screen on the console would go blank.The clinician also reported a lot of beeping at the time of the event.Troubleshooting was performed by turning the console off and restarting it after the first error.Rotations per minute were unable to be increased and the machine constantly beeped.The machine was unable to be silenced, then an error message of "motor failure" came across followed by the console screen going black.The clinician then got a new console and plugged in the motor.When it was turned on, the same exact problem persisted.The clinician then got a new motor and plugged it in and tested it, and both consoles functioned as intended which allowed the clinician to confirm that the motor was the issue.
 
Manufacturer Narrative
Section a: no patient was involved.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, US
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18961974
MDR Text Key338387688
Report Number3003306248-2024-00455
Device Sequence Number1
Product Code KFM
UDI-Device Identifier07640135140078
UDI-Public07640135140078
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number102956
Device Lot Number7316527
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-