It was reported that during a patient visit the system controller was noted to have a driveline power fault, as well as dust/debris in the modular cable connection.The modular cable and controller were exchanged and no further alarms were noted.Log files were submitted for review and contained the onset of driveline power a faults on (b)(6)2024.There were also 2 separate sets of alarms associated with driveline disconnects on (b)(6)2024.It was noted that following the disconnects the driveline fault appeared to have persisted.The driveline disconnects were reportedly not related to troubleshooting, and it was unknown if the modular cable had any trauma, though no visible damage was noted.Related manufacturer's reference number for the pump: 2916596-2024-01494 related manufacturer's reference number for the modular cable: 2916596-2024-01496.
|