Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6 investigation summary the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returning device found that the depth gauge for 2.0mm and 2.4mm screws was broken from the needle, the broken fragment was returned for evaluation.A dimensional inspection was not performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.Additionally the protection sleeve was missing.A root cause for this condition cannot be determined.The overall complaint was confirmed as the observed condition of the depth gauge for 2.0mm and 2.4mm screws would contribute to the device issue.There is no indication that a design or manufacturing issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Dimensional inspection: n/a device history part # 319.006 synthes lot # h521672 supplier lot # h521672 release to warehouse date: 19 nov 2018 supplier : (b)(4).No ncrs were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|