Catalog Number D134701 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient underwent a cardiac ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and the ablation catheter could not be flushed.The ¿tubing set ventil upstream the catheter¿ was opened, and the flow was correct.As the ¿ventil was opened direction catheter¿, the flow was blocked.There was no error from the pump.Correct catheter settings were selected on the device.The catheter was changed, and the issue was resolved.No consequences for the patient and no delay.
|
|
Manufacturer Narrative
|
E 1.Initial reporter phone : (b)(6).If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
|
Manufacturer Narrative
|
The biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
Search Alerts/Recalls
|
|