MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72404250
Device Problems Fluid/Blood Leak (1250); Mechanical Problem (1384); Material Puncture/Hole (1504); Inadequacy of Device Shape and/or Size (1583); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/05/2024
Event Type  Injury  
Event Description
It was reported that this inflatable penile prosthesis stopped working as the patient could not pump the device.A surgical procedure was performed, during which fluid loss caused by a hole in the cylinder tubing was identified.In addition, it was noted that the implant was significantly smaller than what was measured.All components were removed and replaced.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18966234
MDR Text Key338431893
Report Number2124215-2024-17017
Device Sequence Number1
Product Code FHW
UDI-Device Identifier00878953003580
UDI-Public00878953003580
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
N970012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/01/2016
Device Model Number72404250
Device Catalogue Number72404250
Device Lot Number891899002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/18/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
RESERVOIR: UPN: 72404155 / LOT: 906108002
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age76 YR
Patient SexMale
-
-