• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported a perforator (id 261221) drill disassembled when pulling the device from cranium after making the final burr hole.The inner drill remained in the cranium, so the drill was removed using pliers.Total number of burr hole made by the product is unknown.The drill used with the perforator is unknown.No dural damage occurred to the patient.It was stated that there was a possibility of the product remained in the patient's body.According to information provided, it is unknown if there was a surgical delay due to drill disassemble.It is also unknown if all broken parts were recovered and if x-ray was taken to assure no parts left in patient.It is unknown if the drill is electric or pneumatic, if the perforator clicked in place in the drill, and if the recommended spring tests being performed between each burr hole.
 
Manufacturer Narrative
Additional information received: - the drill used with the perforator: anspach - all the broken parts were recovered.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18966514
MDR Text Key339261215
Report Number3014334038-2024-00056
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number261221
Device Lot Number7283203
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/01/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
DRILL - ANSPACH.
-
-