Model Number TJF-Q180V
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  malfunction  
Event Description
It was observed that during the device inspection, the duodenoscope exhibited that the distal end was cracked.There were no reports of patient involvement.
 
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the investigation, it is likely that ingredients of glue were deteriorated by chemical solutions used during reprocessing.However, a definitive root cause could not be determined.The instruction manual states the detection method associated with the event in " 3.2 inspection of the endoscope: inspection of the forceps elevator mechanism: inspection for smooth operation: visually confirm that the portion of the elevator wire extending from the distal end of the endoscope is not broken, frayed, or bent (see figure 3.5).If any damage (broken, frayed, or bent portion) is observed on the elevator wire as shown in figure 3.5, do not use the endoscope".Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18966596
MDR Text Key338700041
Report Number9610595-2024-06244
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170367311
UDI-Public04953170367311
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K143153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberTJF-Q180V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Was the Report Sent to FDA? No
Date Manufacturer Received04/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-