The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the electrical pin connector fluid damage, the lg connector fluid damage, the segment broken, the lcb (light carrying bundle) broken, the insertion flexible tube buckled, the light guide cable buckled, the control body corroded, the operation channel (primary) leak, the root brace rubber (lg control body) cut, the objective lens scratched, and the u/d and the r/l pulley wires worn out; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|