PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MULTI MEASUREMENT SERVER X2 Back to Search Results
Model Number M3002A
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/13/2024
Event Type  malfunction  
Event Description
The customer reported that there was a technical alarm message and no more sound on the x2 connected to a monitor.It is unknown if the device was in use at time of event, and there was no adverse event reported.The problem was isolated at the level of the x2; the welding of the loudspeaker is detached.A replacement speaker was ordered.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.Corrected data: e1: reporter institution phone number (b)(6).E1: reporter phone number (b)(6).
 
Manufacturer Narrative
Philips received a complaint on the intellivue multi measurement server x2 indicating that there was a loudspeaker fault inop and no more sound.A remote service engineer (rse) spoke to the customer and determined that the speaker required replacement.Based on the information available and the testing conducted, the cause of the reported problem was the speaker.The reported problem was confirmed.The customer was provided a replacement speaker to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MULTI MEASUREMENT SERVER X2
Type of Device
INTELLIVUE MULTI MEASUREMENT SERVER X2
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key18968509
MDR Text Key338609983
Report Number9610816-2024-00168
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838000261
UDI-Public00884838000261
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K150310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3002A
Device Catalogue NumberM3002A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-