Model Number LSP202V
Device Problems Failure to Capture (1081); Failure to Sense (1559); Device Dislodged or Dislocated (2923)
Patient Problem Embolism/Embolus (4438)
Event Date 03/06/2024
Event Type  Injury  
Event Description
It was reported that at discharge check, the patient's right ventricular leadless pacemaker (rv lp) exhibited failure to sense and failure to capture.Chest x-ray revealed the rv lp had dislodged into the hepatic vein.The device was successfully retrieved and replaced.The patient was stable.
 
Manufacturer Narrative
The reported events of post-op dislodgement, failure to capture and failure to sense could not be confirmed.Final analysis found the helix length was within specification.A device history record (dhr) review was performed.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.No anomalies were detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVEIR¿ LEADLESS PACEMAKER, RIGHT VENTRICLE
Type of Device
Leadless pacemaker
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18969224
MDR Text Key338496709
Report Number2017865-2024-35791
Device Sequence Number1
Product Code PNJ
UDI-Device Identifier05415067040725
UDI-Public05415067040725
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P150035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLSP202V
Device Lot NumberS000091824
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/06/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ATRIAL LEADLESS PACEMAKER
Patient Outcome(s) Required Intervention;
Patient Age88 YR
Patient SexFemale
-
-