It was reported that explant indicator was reached on (b)(6) 2020 for this patient with an implantable cardioverter defibrillator (icd).Technical services reviewed device data and found there was a code 1007.The last cap reform measured 10.2 seconds.It was recommended to check the device with a programmer.Disk analysis was performed and found the end of life (eol) was triggered by the failure to charge.The cause of the long charge times is unknown.It was noted that the patient was in hospice and a magnet was placed over the device.It was noted the passed away however, the death is not related to the code 1007.The device most likely will not be returned.No adverse patient effects were reported.
|