Model Number 0292
Device Problem High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2024
Event Type  malfunction  
Event Description
It was reported that this right ventricular (rv) lead exhibited high out of range shock impedance measurements.Data analysis was performed, and it was confirmed that for the past year, the shock impedance has gradually increased, recently reaching values greater than 125 ohms, but the pace-sense impedance and r-wave amplitude measurements have been stable and within range, and the stored events do not show any evidence of noise.Technical services (ts) did not find any evidence that could point towards lead impairment, but troubleshooting steps were provided to assess this products integrity, including electrical measurements, patient maneuvers, x-ray imaging and commanded shock testing.At this time, no further information is available.The lead remains in service and no adverse patient effects have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE SG
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18969505
MDR Text Key338665122
Report Number2124215-2024-17839
Device Sequence Number1
Product Code NVY
Combination Product (y/n)Y
Reporter Country CodeUK
PMA/PMN Number
P910073/S077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/20/2014
Device Model Number0292
Device Catalogue Number0292
Device Lot Number115266
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/20/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial