It was reported that this right ventricular (rv) lead exhibited high out of range shock impedance measurements.Data analysis was performed, and it was confirmed that for the past year, the shock impedance has gradually increased, recently reaching values greater than 125 ohms, but the pace-sense impedance and r-wave amplitude measurements have been stable and within range, and the stored events do not show any evidence of noise.Technical services (ts) did not find any evidence that could point towards lead impairment, but troubleshooting steps were provided to assess this products integrity, including electrical measurements, patient maneuvers, x-ray imaging and commanded shock testing.At this time, no further information is available.The lead remains in service and no adverse patient effects have been reported.
|