• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL X1 LARGE JAW TISSUE SEALER; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL X1 LARGE JAW TISSUE SEALER; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number NSLX120L
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 3/25/2024 b3: event year only reported: 2024 d4 batch #: x70344 attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during an unknown procedure the device clamped on the blood vessel properly but the jaws wouldn't open up when trying to remove.They had to carefully slide the jaws off the vessel.No patient harm reported.No additional information provided.
 
Manufacturer Narrative
(b)(4).Date sent: 4/16/2024.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with no apparent damage.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All tones were heard during functional testing.No issues were noted with the opening and closing of the jaws.There were no anomalies noted with the functionality of the device.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL X1 LARGE JAW TISSUE SEALER
Type of Device
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
JABIL (TAJUANA)
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18969919
MDR Text Key338725846
Report Number3005075853-2024-02367
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10705036015024
UDI-Public10705036015024
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNSLX120L
Was Device Available for Evaluation? No
Date Manufacturer Received03/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GENERATOR
-
-