MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
|
Back to Search Results |
|
Model Number BI70002000 |
Device Problems
Image Display Error/Artifact (1304); Poor Quality Image (1408)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/20/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H3, h6: the system was serviced in the field by a medtronic representative.The rep completed fluoro gain calibration and the issue resolved.Codes b01, c13, and d02 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used in an unknown procedure.It was reported that after taking images, there was a white circle around the area of interest.The site said that the images were washed out with symmetrical striations on one side.The site indicated that they had completed the rad gain calibrations "in the last few days." no delay or patient impact information was provided.
|
|
Manufacturer Narrative
|
H2: g2 has been updated to correctly reflect the report source for this event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
Concomitant medical product: section d information references the main component of the system.Other relevant device(s) are: product id bi71000901 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|