• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN SIGNIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN SIGNIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIGPHANDLE
Device Problems Retraction Problem (1536); Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Event Description
According to the reporter, during a laparoscopic pancreatic tail resection, the firing stopped midway.The firing was repeated in advance by one cm and waited for about a minute.When the knife reaches the tip, a power handle error with a red circle; a slash and an exclamation mark appeared and the knife did not return.The surgeon pressed and held the force reset button to return the knife and release it from the tissue.When the sales rep checked the handle after the surgery, it was confirmed that the handle attaching a new shell and reload was able to be used with sponge successfully.A new powered stapler and shell were used to resolve the issue.No patient injury.
 
Manufacturer Narrative
D10 concomitant products: sigpshell, sig power sigpshell control shell (lot#unknown); sigadaptstnd, sig power sigadaptstnd linear adapter (sn:unknown); sigtrsb60axt, sigtrsb60axt 60mm xtr thk reinforced rel (lot#unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGNIA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18970384
MDR Text Key339116285
Report Number1219930-2024-01308
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884521543782
UDI-Public10884521543782
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K160176
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSIGPHANDLE
Device Catalogue NumberSIGPHANDLE
Was Device Available for Evaluation? No
Date Manufacturer Received02/29/2024
Date Device Manufactured02/01/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
SEE H11.
-
-