MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 0292
Device Problems High impedance (1291); Material Integrity Problem (2978); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2024
Event Type  Injury  
Event Description
It was reported that this right ventricular (rv) defibrillation lead exhibited high threshold measurements and high out of range pacing impedance measurements greater than 2,000 ohms.Surgical intervention was undertaken.The lead was explanted and sustained some damage during the explant procedure.The patient was then implanted with a subcutaneous implantable cardioverter defibrillator (s-icd) system.No additional adverse patient effects were reported.The product is expected to be returned for analysis.This report will be updated upon return and completion of analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE SG
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18972302
MDR Text Key338517872
Report Number2124215-2024-17906
Device Sequence Number1
Product Code NVY
Combination Product (y/n)Y
Reporter Country CodeNZ
PMA/PMN Number
P910073/S077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2012
Device Model Number0292
Device Catalogue Number0292
Device Lot Number012299
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/30/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-