BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number R5C4482 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
E1: initial reporter phone no.(b)(6).The device was received for evaluation.A visual inspection with the naked eye noted a white twist clamp would not fully align with the notch on the main body.Functional tests including leak, clear passage, and clamp function tests were performed.No issue was noted during the leak and clear passage testing.Clamp function tests identified a clamp closed not allowing flow through closed clamp.The twist clamp detent would not fully engage.The reported condition was verified.The cause of the condition was related to the milling process during manufacturing.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
During evaluation, the white twist clamp would not fully align with the notch on the main body was identified on the minicap transfer set.There was no patient involvement.No additional information is available.
|
|