• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 37612
Device Problems Failure to Charge (1085); Material Frayed (1262); Failure to Power Up (1476); Charging Problem (2892); Connection Problem (2900); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2022
Event Type  malfunction  
Manufacturer Narrative
Continuation of d10: product id 37761 ,serial# (b)(6), product type recharger.Product id 37751 lot# serial# (b)(6), product type recharger.Section d information references the main component of the system.Other relevant device(s) are: product id: 37761, serial/lot #: (b)(6).Product id: 37751, serial/lot #: (b)(6).Analysis of the recharger (b)(6) revealed that the antenna had a frayed cord and the connector pin was bent.Analysis of the desktop charger (b)(6) revealed that the connector pin was bent.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that the recharger was not charging and reported this morning the recharger wouldn't power on.The caller confirmed no visible damage to the recharger port.The patient (pt) stated the green light of the ac power supply was on and there was no visible damage to the desktop charger (dtc) cord, but that the connector pin of the dtc was bent.The issue was not resolved.An email was sent to the repair department to replace the dtc.No symptoms were reported.Additional information was received from the consumer who reported they received the replacement dtc but the recharger still wasn't charging like normal and they also discovered that the recharger antenna was frayed and recharger was showing a 376 error code.Email sent to repair to send replacement recharger.No symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18975198
MDR Text Key338592406
Report Number3004209178-2024-07768
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00643169864238
UDI-Public00643169864238
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2019
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2022
Date Manufacturer Received06/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR