Additional information: d4, d9, h3, h4, h6, h10.H4: the lot was manufactured between july 7, 2023 and july 11, 2023.H10: the device was received for evaluation.During visual inspection fluid was observed inside the bag that contained the unit suggesting a leak may have occurred.When the unit was removed from the bag, the cause of leak was found to be a partially broken tubing at the junction of the luer.The reported condition was verified.The cause of the condition could not be determined; however, the most probable cause may be due to how the device was being handled during use.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|