Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Section e.1: the initial reporter phone: (b)(6).The initial reporter email address was not available / reported.Based on complaint information, the device remains implanted and is thus not available for evaluation.Lake region medical performed a review of the device history records relative to the manufacturing, inspection, and packaging of the lot 7773924.The history record indicates this product was final inspection tested at lake region medical and was determined to be acceptable.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.The incomplete expansion of the enterprise2 stent can potentially lead to thrombosis and/or migration or embolization, resulting in ischemia or infarct.In this case, the treating physician was required to implant a second stent inside the enterprise2 stent to open the stent completely and preclude patient harm.Based on this required surgical intervention/surgical procedure modification, this event meets us fda reporting criteria under 21 cfr 803 with a classification of ¿serious injury.¿ the file will be re-reviewed if additional information is received at a later date.With the information available and without the complaint product available to be returned for analysis, the reported product issue documented in the complaint cannot be confirmed through functional evaluation and analysis.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint device; however, it is possible that clinical and procedural factors, including device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|