Model Number 37612
Device Problems Material Frayed (1262); Failure to Power Up (1476); Charging Problem (2892); Connection Problem (2900); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2021
Event Type  malfunction  
Event Description
It was reported that the recharger (insr) does not work anymore.Insr won't power on or charge from desktop charger (dtc).The cord leading dtc was frayed, connector pin was bent and the inside of the insr port was damaged.The insr antenna cord was frayed.An email was sent to the repair department to replace the dtc and insr.
 
Manufacturer Narrative
Continuation of d10: product id 37651, serial# (b)(6).Product type recharger product id 37761, serial# (b)(6).Product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: 37651, serial/lot #: (b)(6), udi#: (b)(4); product id: 37761, serial/lot #: (b)(6).H3: analysis of the recharger (serial #:(b)(6)) revealed no anomalies with the device.Analysis of the recharger (serial #: (b)(6)) revealed the connector pin was bent.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18976047
MDR Text Key338573245
Report Number3004209178-2024-07781
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00643169864238
UDI-Public00643169864238
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2020
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2021
Date Manufacturer Received10/24/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H11...."
Patient Age73 YR
Patient SexMale
-
-