Model Number 3501
Device Problems Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 03/19/2024
Event Type  malfunction  
Event Description
It was reported that the subcutaneous implantable cardioverter defibrillator (s-icd) delivered an inappropriate shock due to t-wave oversensing and small signal amplitudes.The s-icd was reprogrammed from the secondary to alternate sensing vector.The s-icd system remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18976147
MDR Text Key338664887
Report Number2124215-2024-18057
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526599200
UDI-Public00802526599200
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110042/S077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/23/2024
Device Model Number3501
Device Catalogue Number3501
Device Lot Number215087
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age42 YR
Patient SexMale
-
-