It was reported that the patient presented to the hospital for a scheduled implant procedure.During the procedure, the left ventricular (lv) lead was found difficult to be separated from the guidewire.Upon removal of the guidewire from the lv lead, the distal tip of the guidewire was found broken.The physician elected to remove and replace the lv lead as a precautionary measure on (b)(6) 2024.The patient was in stable condition throughout.
|
The reported event difficult to remove stylet/guidewire was confirmed.As received, a complete lead was returned in one piece.Visual inspection and x-ray examination to the distal region and the entire lead found no anomalies, but the inner coil was bunched up at the connector region, consistent with procedure damage.Electrical testing found no conductor fractures or internal shorts.The cause of the reported event was due to inner coil damage due to procedural damage.No other anomalies were seen.
|