Model Number CD3369-40C
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pocket Erosion (2013)
Event Date 03/04/2024
Event Type  Injury  
Event Description
It was reported that the patient presented in the clinic with pocket erosion.The implantable cardioverter-defibrillator (icd) was explanted and replaced to resolve the issue.The patient was in stable condition throughout the procedure.
 
Manufacturer Narrative
Further information requested but was not received.
 
Manufacturer Narrative
The reported event was pocket erosion.The device was returned for analysis.Interrogation results were normal, visual screen was acceptable, and preliminary test results were acceptable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18976705
MDR Text Key338548225
Report Number2017865-2024-36886
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508346
UDI-Public05414734508346
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Model NumberCD3369-40C
Device Lot NumberA000117199
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUARTET LEAD.
Patient Outcome(s) Required Intervention;
Patient Age64 YR