Event Description
It was reported that during an angioplasty procedure, when the pressure was pumped to 8 atm during repeat dilatation, pressure pump abnormality was noticed and blood was allegedly seen on gentle retraction.It was further reported that the pta balloon allegedly could not be fully retracted.Furthermore, upon removal of the balloon from the sheath, the balloon was allegedly found to be broken.The procedure was completed by using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a video was provided for review.The investigation of the reported event is currently underway.H10: d4.(expiration date: 10/2025).H11: section a: through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18976712
MDR Text Key338671286
Report Number2020394-2024-00391
Device Sequence Number1
Product Code DQY
UDI-Device Identifier00801741062872
UDI-Public(01)00801741062872
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K083657
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCQ5064
Device Lot NumberREGY0978
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-