Model Number MC1AVR1 |
Device Problems
Migration or Expulsion of Device (1395); Pacing Problem (1439); Capturing Problem (2891); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/19/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was reported that approximately two days post implant, the leadless implantable pulse generator (ipg) exhibited loss of pacing due to rising and high thresholds.A micro dislodgement was believed to be the case.The leadless ipg was reprogrammed and remains in use but pacing is still not possible. no patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was further noted that the leadless ipg was explanted and replaced by an ipg system.
|
|