Model Number MI1250 SYNCHRONY 2 |
Device Problem
Migration (4003)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 02/16/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
During a revision operation for a migration of the array, it was noted that the contacts were sheared off.Therefore the device was replaced.
|
|
Event Description
|
During a revision surgery for a migration of the array, it was noted that the contacts were sheared off.Therefore the device was replaced.
|
|
Manufacturer Narrative
|
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the revision surgery.According to the information received, the device was not providing satisfactory benefit to the recipient due to a post-operative active electrode migration out of the cochlea.A revision surgery was planned, however, during revision surgery the device was damaged and therefore explanted.This is a final report.
|
|