• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SYNCHRONY; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SYNCHRONY; COCHLEAR IMPLANT Back to Search Results
Model Number MI1250 SYNCHRONY 2
Device Problem Migration (4003)
Patient Problem Failure of Implant (1924)
Event Date 02/16/2024
Event Type  malfunction  
Manufacturer Narrative
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
During a revision operation for a migration of the array, it was noted that the contacts were sheared off.Therefore the device was replaced.
 
Event Description
During a revision surgery for a migration of the array, it was noted that the contacts were sheared off.Therefore the device was replaced.
 
Manufacturer Narrative
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the revision surgery.According to the information received, the device was not providing satisfactory benefit to the recipient due to a post-operative active electrode migration out of the cochlea.A revision surgery was planned, however, during revision surgery the device was damaged and therefore explanted.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key18977774
MDR Text Key338677000
Report Number9710014-2024-00282
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09008737366827
UDI-Public(01)09008737366827
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/19/2023
Device Model NumberMI1250 SYNCHRONY 2
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/26/2024
Date Device Manufactured04/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age52 YR
Patient SexFemale
-
-