SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Model Number HL-90 |
Device Problem
Failure to Pump (1502)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/28/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the pump failed.The issue was not related to damage/abuse, nor was the unit dropped.There was no patient involvement and no patient harm/adverse event reported.
|
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection found: enclosure cracked under quick connect, corroded quick connect, and outdated quick connect, pcb, and power switch.Functional testing confirmed the complaint.Root cause was attributed to a faulty pump.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
|
|