• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER420
Device Problem Failure to Fire (2610)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 3/26/2024.D4: batch # x7033v.Investigation summary: the product was returned to ethicon  for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the er420 device was returned with a clip in the jaws; the clip was removed in order to inspect the jaws and they were found to be misaligned.In an attempt to replicate the reported incident, the device was tested for functionality; during the analysis, the instrument was cycled and it fed and formed eleven(11) scissored clips due to the misaligned condition of the jaws; finally the instrument locked out as intended.The damage in the device is related to improper use of the device.Possible causes for the condition found may be if the device is closed over an existing hard object or clip placing stress on the jaws causing them to distort or yield and not return to their original dimensions/position or excessive application of torque to the jaws when positioning the device on a vessel.The reported complaint was confirmed.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Event Description
It was reported that during a laparoscopic cholecystectomy, clip jamming occurred at 3rd firing.The surgeon did not twist the device.The device was used on the gallbladder artery.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18978437
MDR Text Key338615696
Report Number3005075853-2024-02404
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012597
UDI-Public10705036012597
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberER420
Device Lot NumberX70352
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2024
Date Manufacturer Received03/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/15/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-