• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Device Problems Failure to Fire (2610); Mechanical Jam (2983); Firing Problem (4011)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 3/26/2024.D4: batch # a9e78y.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported event, the device was tested for functionality.Upon testing, the device was fired, and the clips did not advance into the jaw.The device was noted to have the tip of the advancer bent.The instrument was disassembled, upon disassembly the advancer was confirmed to be bent and five(5) clips were found inside clip track.However, it is known from the history of the device that bent advancer condition may lead to malformed clips.The event reported was confirmed and it is related to improper use of the device.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.Please reference the instruction for use for more information.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Event Description
It was reported that during an unknown procedure, the applier stopped reloading the clips in the jaw and could not be fired.It also took more effort than normal to close the applier.No patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18978504
MDR Text Key338597059
Report Number3005075853-2024-02405
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEL5ML
Device Lot NumberA9E78Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2024
Date Manufacturer Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-