Model Number 6947M55 |
Device Problems
High impedance (1291); Impedance Problem (2950)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was reported that the right ventricular (rv) coil had rising impedance, an impedance spike, and high impedance.Additionally the superior vena cava (svc) coil had rising impedance, an impedance spike, and high impedance.The lead was capped and replaced.No patient complications have been reported as a result of this event.
|
|
Search Alerts/Recalls
|
|