It was reported that the patient experienced discomfort.During the remote follow up, loss of capture and loss of sensing was noted on the right ventricular (rv) lead due to dislodgment.The device was reprogrammed to disable high voltage therapy and the patient was hospitalized.A revision procedure was performed, however, while repositioning the rv lead, the helix was unable to extend.The rv lead was explanted and replaced to resolve the event.The patient was in stable condition.
|
The reported events were lead dislodgement, loss of capture, loss of sensing, and helix mechanism anomaly.As received, a complete lead was returned in one piece.The reported event of helix mechanism anomaly was confirmed.Visual examination of the lead found the helix was retracted and clogged with blood/tissue.X-ray examination of the lead found the inner coil in the connector region was over torqued consistent with procedural damage.After cutting the lead, cleaning the distal portion, and applying torque directly to the inner coil, the helix was able to extend and retract.The measured full helix extension length was within specification.The cause of the reported event of helix mechanism anomaly was due to the over torqued inner coil in the connector region and the helix clogged with blood/tissue.The reported events of loss of capture and loss of sensing were not confirmed.Visual and x-ray examination of the lead did not find any anomalies with the exception of procedural damage.Electrical testing did not find any indication of conductor fractures or internal shorts.
|