Model Number A219
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 03/06/2024
Event Type  malfunction  
Event Description
It was reported that, this subcutaneous implantable cardioverter defibrillator (s-icd) delivered an inappropriate shock due to low r wave and t wave and myopotentials oversensing.The patient was seeing in the clinic and noted the shock was received getting out of the bath.An auto setup was performed, and the device chose primary sensing vector.The field representative stressed out primary with manoeuvres and the highly noisy captured in the secg was the patient doing pushups.X rays were performed, and the system seemed higher.This s-icd remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM MRI S-ICD
Type of Device
SUBCUTANEOUS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (S-ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18979836
MDR Text Key338760257
Report Number2124215-2024-18218
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberA219
Device Catalogue NumberA219
Device Lot Number186888
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-