BOSTON SCIENTIFIC CORPORATION AUTOGEN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G179 |
Device Problems
Premature Discharge of Battery (1057); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/08/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that this pacemaker was suspected of exhibiting premature battery depletion behavior.During a routine follow-up visit, the estimated battery longevity was showing only 1.5 years remaining with 500% power consumption.The physician requested that the device be re-interrogated to determine if there were any optimizations needed to assist with battery longevity.Another follow-up with the patient was scheduled a few days later after the previous follow-up visit.This time, the battery longevity was showing 1 year remaining and the power consumption increased to 536% usage.During this interrogation, all testing was completed, and the values all appeared within normal range.Device settings were evaluated with all settings seeming optimized.A copy of device memory was saved and submitted to technical services (ts) for further review.A ts consultant confirmed the suspected premature depletion and recommended immediate replacement.This device currently remains implanted and in service.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that this pacemaker was suspected of exhibiting premature battery depletion behavior.During a routine follow-up visit, the estimated battery longevity was showing only 1.5 years remaining with 500% power consumption.The physician requested that the device be re-interrogated to determine if there were any optimizations needed to assist with battery longevity.Another follow-up with the patient was scheduled a few days later after the previous follow-up visit.This time, the battery longevity was showing 1 year remaining and the power consumption increased to 536% usage.During this interrogation, all testing was completed, and the values all appeared within normal range.Device settings were evaluated with all settings seeming optimized.A copy of device memory was saved and submitted to technical services (ts) for further review.A ts consultant confirmed the suspected premature depletion and recommended immediate replacement.This device currently remains implanted and in service.No adverse patient effects were reported.Additional information: new information was provided from the field representative indicating that this device was explanted and replaced.No additional adverse patient effects were reported.This device is expected for return.
|
|
Manufacturer Narrative
|
This device is expected for return.Following completion of laboratory analysis, this event will be further updated.
|
|
Search Alerts/Recalls
|
|
|