Model Number 560BCS1
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Event Description
During preventive maintenance by a service technician this bio-console 560 controller instrument had an mpu board which had a hard offset alarm.This was detected during service so there was no patient involvement, so no adverse effect occurred.
 
Manufacturer Narrative
Device evaluation summary: during preventive maintenance by service technician, it was observed that bio-console 560 had a mpu board issue, with a hard offset alarm.Issue was resolved by replacing the mpu board.Preventive maintenance was performed as per specification.D9: instrument was serviced at the facility by medtronic field service and was not returned to medtronic service center medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO CONSOLE 560
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18980101
MDR Text Key338620072
Report Number2184009-2024-00132
Device Sequence Number1
Product Code DWA
UDI-Device Identifier00763000236526
UDI-Public00763000236526
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080824
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number560BCS1
Device Catalogue Number560BCS1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-