• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT IRELAND DIAGNOSTICS DIVISION ARCHITECT FREE T4 REAGENT KIT; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND DIAGNOSTICS DIVISION ARCHITECT FREE T4 REAGENT KIT; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 07K65-77
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/20/2024
Event Type  malfunction  
Manufacturer Narrative
All available patient information was included.Additional patient details are not available.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer reported a falsely elevated architect free t4 result for one patient sample when comparing the initial result to the retest result.The following data was provided (reference range: 9.01-19.05 pmol/l or 0.70-1.48 ng/dl): initial free t4 result = > 64 pmol/l (> 5.00 ng/dl) repeat free t4 result after centrifugation = 19.05 pmol/l (1.48 ng/dl) there was no impact to patient management reported.
 
Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history record review, and in-house testing was performed on lot 53725ud02.Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.A search for similar complaints did identify an increase in complaint activity for lot 53725ud02, however, no related trends were identified regarding commonalities for complaint lot number and issue.Additionally, in-house performance testing was completed which indicates the product is performing as expected.A review of the device history record did not identify any potential non-conformances, non-conformances, or deviations associated with the complaint issue.Labeling was reviewed and adequately addresses the issue under review.Based on the investigation, no systemic issue or product deficiency of the architect free t4 reagent lot 53725ud02 was identified.
 
Event Description
The customer reported a falsely elevated architect free t4 result for one patient sample when comparing the initial result to the retest result.The following data was provided (reference range: 9.01-19.05 pmol/l or 0.70-1.48 ng/dl): initial free t4 result = > 64 pmol/l (> 5.00 ng/dl); repeat free t4 result after centrifugation = 19.05 pmol/l (1.48 ng/dl).There was no impact to patient management reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT FREE T4 REAGENT KIT
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18980208
MDR Text Key338620948
Report Number3005094123-2024-00140
Device Sequence Number1
Product Code CEC
UDI-Device Identifier00380740173678
UDI-Public(01)00380740173678(17)240629(10)53725UD02
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K173122
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07K65-77
Device Lot Number53725UD02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/29/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARC I2000SR INST, 03M74-02, (B)(6); ARC I2000SR INST, 03M74-02, (B)(6)
-
-