It was reported that this patient with this left ventricular (lv) lead and cardiac resynchronization therapy defibrillator (crt-d) experienced ventricular fibrillation (vf) in which undersensing was observed.This patient has a left ventricular assist device (lvad) that was possibly being oversensed by this lv lead.The field representative noted there was an episode the same day with ventricular tachycardia (vt) which was successfully converted with anti-tachycardia pacing (atp).Technical services (ts) discussed algorithms and programming with the field representative along with suggesting reviewing with the electrophysiology to confirm if they want to adjust sensitivity, rate count or adjust this lv lead position.Ts noted this may be limited due to low r wave amplitude, patient condition lvad and oversensing of lvad.This lv lead remains in service.No adverse patient effects were reported.
|