Model Number 6935M72 |
Device Problems
Capturing Problem (2891); Impedance Problem (2950)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2024 |
Event Type
Injury
|
Event Description
|
It was reported that thirteen days post implant that the right ventricular (rv) lead had increased pacing impedance and thresholds.The lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Continuation of d10: dvpb3d4 icd implanted (b)(6) 2020 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
Product event summary: the full lead was returned and analyzed.Analysis indicated the setscrew marks on the connector pin were too proximal.Visual analysis of the lead indicated damage at implant.The analyst noted there is only one set of setscrew marks on the df-4 pin and it is too proximal - lead was not fully inserted into the device during implant.All electrical testing was within specified parameters.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|