• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0706-XTW
Device Problem Incomplete Coaptation (2507)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 03/04/2024
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.All available information was investigated and the reported slda per the physician is related to friable and tethered leaflets and is therefore, related to patient conditions.Unspecified tissue injury appears to be due to the slda.Tissue injury/damage is listed in the instructions for use (ifu) as a known possible complication associated with mitraclip procedures.Serious injury/ illness/ impairment was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Event Description
It was reported this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4, restricted posterior, and thick leaflets.An xtw clip (31024a2072) was inserted, but upon closing the clip, a tear on the posterior leaflet was suspected.The clip was then repositioned and deployed.To further reduce mr, an additional xtw clip (31107r1082) was inserted and deployed.However, after deployment, the clip detached from posterior leaflet and remained attached to the anterior leaflet (single leaflet device attachment/slda).It was observed that there was damage to the posterior leaflet.Upon further review, the first clip had detached from the anterior leaflet and remained attached to the posterior leaflet (slda).No additional clips were implanted, and mr remained at a grade of 4.There was no clinically significant delay in the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18980819
MDR Text Key338660878
Report Number2135147-2024-01352
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037381
UDI-Public(01)05415067037381(17)241105(10)31107R1082
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0706-XTW
Device Lot Number31107R1082
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/04/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANTED MITRACLIP; STEERABLE GUIDE CATHETER
Patient Outcome(s) Other;
Patient Age72 YR
Patient SexFemale
Patient Weight71 KG
-
-