• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2024
Event Type  Injury  
Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that when using the light source, during an unknown therapeutic procedure the image suddenly blurred out without any error messages on the screen.Troubleshooting to change the camera head and telescope was not successful.The product was inspected, and no abnormalities were detected.The procedure was delayed for more than 30 minutes.The anesthesia was not extended, and procedure was completed with a similar device.There no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Additionally, to provide a correction to the initial with information inadvertently left out g2 and to provide an update to field h4.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the subject device was not returned and a specific root cause of the blurry image that led to a 30-minute delay could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18981851
MDR Text Key338635001
Report Number3002808148-2024-02890
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/17/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-