BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; PACEMAKER Back to Search Results
Model Number L331
Device Problems Pacing Problem (1439); Under-Sensing (1661)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Event Description
It was reported that this pacemaker was undersensing atrial activity, some p waves where falling into post-ventricular atrial refractory period (pvarp) which caused the pacemaker to not pace in the ventricle when required.Technical services (ts) was consulted and provided reprogramming options.This pacemaker remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI EL DR
Type of Device
PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18982448
MDR Text Key338760281
Report Number2124215-2024-18300
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526559266
UDI-Public00802526559266
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberL331
Device Catalogue NumberL331
Device Lot Number611365
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/28/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age34 YR
Patient SexFemale
-
-