It was reported that on (b)(6) 2023 , the procedure was performed to treat a 70% stenosed lesion in the left circumflex (lcx) artery.The bmw elite guidewire (gw) was used in the procedure; however, the tip of the gw broke in the patient.A stent was implanted to embed the separated tip of the gw.A month after the procedure, the patient experienced a headache and sought treatment.It was found that the tip of the gw had migrated to the intracranial blood vessels of the patient.Neurological surgery was performed to remove the tip of the gw from the patient.No additional information was provided.
|
The incident information was reviewed; however, the product was not returned for analysis.The device history record (dhr) and exception review was performed and revealed no indication of a product quality issue.A review of the electronic lot history record and similar complaint query was not performed because the part/lot numbers were not reported.The investigation was unable to determine a conclusive cause for the reported material separation; however, the subsequent patient effects and treatments appear to be related to the operational context of the procedure.Factors that may contribute to tip separation during use include, but are not limited to, interaction with accessory devices, interaction with the patient¿s anatomy, damage incurred during preparation, or excessive force.Based on the information provided and because the device was not returned for analysis, it is unknown what may have caused the reported tip separation.The reported patient effect of embolism is listed in the hi torque guide wires instructions for use as a known patient effect(s) of coronary stenting procedures.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|