SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Catalog Number HL-390 |
Device Problems
Display or Visual Feedback Problem (1184); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the light emitting diode (led) is stuck at 00 degrees.The fault occurred during testing.There was no report of physical damage/mishandling of the device.There was no patient involvement and no patient harm/adverse event reported.
|
|
Manufacturer Narrative
|
Date of event is unknown; no information has been provided to date.Device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
D9: date returned to mfg.: 3/27/2024.H3 and h6.Evaluation codes: updated.One device was received.Per visual inspection, corroded drain fitting and damaged printed circuit board (pcb).When the device was opened several loose parts that were broken off the liquid crystal display (lcd) fell out and when powered up the device only showed "00".After the pcb was replaced, the device functioned properly.Either the pcb was faulty or the damage to the lcd caused the problem.This device has not been serviced in the past, there is no history to review.Replaced the pcb and drain fitting.Performed preventative maintenance (pm).The device passed all functional and delivery tests.
|
|